92 Bio-inert Quaternary Pump User Manual
7 Error Information
Module Error Messages
Initialization Failed
The module failed to initialize successfully within the maximum time window.
A maximum time is assigned for the complete pump-initialization cycle. If the
time is exceeded before initialization is complete, the error message is
generated.
Probable cause Suggested actions
1 Blocked active inlet valve.
Exchange the active inlet valve.
2 Defective pump drive assembly.
Please contact your Agilent service
representative.
3 Defective main board.
Please contact your Agilent service
representative.