136 1260 Nanoflow Pump User Manual
9 Maintenance
Checking and Cleaning the Solvent Inlet Filters
Checking and Cleaning the Solvent Inlet Filters
When If solvent filter is blocked
Parts required Description
Concentrated nitric acid (65 %)
Bidistilled water
Beaker
Preparations Remove the solvent inlet tube from the inlet port of the solvent selection valve or the adapter at the
active inlet valve
WARNING
When opening capillary or tube fittings solvents may leak out.
The handling of toxic and hazardous solvents and reagents can bear health risks.
Please observe appropriate safety procedures (for example, goggles, safety gloves
and protective clothing) as described in the material handling and safety data sheet
supplied by the solvent vendor, especially when toxic or hazardous solvents are
used.
CAUTION
Small particles can permanently block the capillaries and valves of the module.
Damage of the module.
Always filter solvents.
Never use the module without solvent inlet filter.
NOTE
The solvent filters are located on the low-pressure side of the pump. A blocked filter
therefore does not affect the pressure readings of the pump. The pressure readings cannot
be used to check whether the filter is blocked or not.
NOTE
If the filter is in good condition the solvent will freely drip out of the solvent tube (due to
hydrostatic pressure). If the solvent filter is partly blocked only very little solvent will drip
out of the solvent tube.