142 1260 Nanoflow Pump User Manual
9 Maintenance
Exchanging the Outlet Valve Sieve
Exchanging the Outlet Valve Sieve
1 Using a 1/4 inch wrench disconnect the valve capillary from the outlet
valve.
2 Using the 14 mm wrench loosen the valve and remove it from the pump
body.
3 Remove the plastic cap with the gold seal from the outlet valve.
4 Using a pair of tweezers remove the sieve.
5 Place a new sieve into the recess of the outlet valve and replace the cap with
the gold seal.
When Sieve — whenever the pump seals will be exchanged
Valve — if internally leaking
Tools required Wrench 1/4 inch
Wrench 14 mm
Parts required # p/n Description
1 G1312-60067 Outlet valve, complete
1 5063-6505 Sieve (pack of 10)
NOTE
Before exchanging the outlet valve you can try to clean it in a sonic bath. Remove the gold
seal and the sieve. Place the valve in upright position (onto the plastic cap) in a small
beaker with alcohol. Place in a sonic bath for 5 – 10 min. Insert a new sieve and replace the
gold seal.
NOTE
Check the gold seal. It should be exchanged when strongly deformed. Place the valve in an
upright position, insert the sieve into the recess and replace the gold seal with the cap.