1260 Nanoflow Pump User Manual 153
Maintenance
9
Exchanging the Flow Sensor
Exchanging the Flow Sensor
1 Turn off the pump.
2 Using a 1/4 inch wrench disconnect the capillaries:
• coming from the EMPV.
• going to the injection device (port 1).
3 Unscrew the flow sensor.
4 Re-install the new one.
5 Using a 1/4 inch wrench reconnect the capillaries:
• coming from the EMPV.
• going to the injection device (port 1).
When Extended flow range (100 ul) needed.
Leak on the flow sensor.
Unstable column flow
Flow sensor blocked
Tools required Screwdriver Pozidriv #1
Parts required # p/n Description
1 G1376-60004 Nano Flow Sensor 4 µL (1260 Nanoflow Pump)