Exchanging the Flow Sensor
Exchanging the Flow Sensor
1 Turn off the pump.
2 Using a 1/4 inch wrench disconnect the capillaries:
coming from the EMPV.
going to the injection device (port 1).
3 Unscrew the flow sensor.
4 Re-install the new one.
5 Using a 1/4 inch wrench reconnect the capillaries:
coming from the EMPV.
going to the injection device (port 1).
When Extended flow range (100 ul) needed.
Leak on the flow sensor.
Unstable column flow
Flow sensor blocked
Tools required Screwdriver Pozidriv #1
Parts required # p/n Description
1 G1376-60004 Nano Flow Sensor 4 µL (1260 Nanoflow Pump)