2
Performance Specifications
Analog output For pressure monitoring, 2 mV/bar, one output
Communications Controller-area network (CAN), RS-232C, APG Remote: ready, start, stop and shut-down signals,
LAN optional
Safety and
maintenance
Extensive diagnostics, error detection and display (though Instant Pilot and Data System), leak
detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in
major maintenance areas.
GLP features Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of seal
wear and volume of pumped mobile phase with user-settable limits and feedback messages.
Electronic records of maintenance and errors.
Housing All materials recyclable.
Tabl e 14 Performance Specification Agilent 1260 Infinity Nano Pump (G2226A)
Type Specification