52 1260 Nanoflow Pump User Manual
3 Installing the Module
Connecting Modules and Control Software
Connecting Modules and Control Software
Connecting Modules
1 Place the individual modules in a stack configuration as shown in Figure 8
on page 45.
2 Ensure the power switches on the front of the modules are OFF (switches
stand out).
3 Plug a CAN cable into the CAN connector at the rear of the respective
module (except vacuum degasser).
4 Connect the CAN cable to the CAN connector of the next module, see
Figure 9 on page 46.
5 Press in the power switches to turn on the modules.
Connecting a Vacuum Degasser
1 Place the vacuum degaser in the stack of modules as shown in Figure 8 on
page 45.
2 Ensure the power switch at the front of the vacuum degasser is OFF (switch
stands out).
3 Plug an APG cable into the APG remote connector at the rear of the
degasser.
4 Connect the APG cable to the APG remote connector of the pump, see
Figure 9 on page 46.
5 Press in the power switch to turn on the vacuum degasser.
WARNING
Use of unsupplied cables
Using cables not supplied by Agilent Technologies can lead to damage of the
electronic components or personal injury.
Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.
NOTE
The AUX output is intended for troubleshooting. It provides a DC voltage in the range of 0 –
1 V which is proportional to the vacuum level in the degasser chambers.