1260 Nanoflow Pump User Manual 55
Installing the Module
3
Flow Connections
1 Remove the front cover by pressing the snap fasteners on both sides.
Figure 14 Removing the Front Cover
2 Place the solvent cabinet on top of the module.
3 Remove the sintered glas inlet filters and the filter adapters from the bottle
head assemblies. Replace them by the stainless steel filters from the pump
asseccories kit.
4 Connect the bottle head solvent tubes to the lower ports of the online
degasser. Connect the upper ports of the online degasser to the inlet ports
A1, A2, B1 and B2 of the solvent selection valve of the pump. Fix the solvent
tubes in the clips of pump, degasser and solvent cabinet. Label the solvent
tubings with the provided stick-on labels.
5 Using a piece of emery cloth connect the waste tubing to the EMPV and
place it into your waste system.
6 If the pump is not part of an Agilent 1260 Infinity system stack or placed on
the bottom of a stack, connect the corrugated waste tube to the waste outlet
of the pump leak handling system.
NOTE
Use a piece of sand paper to get a good grip when pushing the stainless steel filters into the
tubings.