92 1260 Nanoflow Pump User Manual
7 Error Information
General Error Messages
Leak Sensor Short
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current increases above the upper limit,
the error message is generated.
Leak Sensor Open
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current falls outside the lower limit, the
error message is generated.
Probable cause Suggested actions
1 Defective flow sensor.
Please contact your Agilent service
representative.
2 Leak sensor incorrectly routed, being
pinched by a metal component.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1 Leak sensor not connected to the main
board.
Please contact your Agilent service
representative.
2 Defective leak sensor.
Please contact your Agilent service
representative.
3 Leak sensor incorrectly routed, being
pinched by a metal component.
Please contact your Agilent service
representative.