1260 Nanoflow Pump User Manual 95
Error Information
7
Module Error Messages
Module Error Messages
Zero Solvent Counter
The error message is triggered if the remaining volume in a solvent bottle falls
below the set limit.
Pressure Above Upper Limit
The system pressure has exceeded the upper pressure limit.
Probable cause Suggested actions
1 Volume in bottle below specified volume.
Refill bottles and reset solvent counters.
2 Incorrect setting.
Make sure the set solvent volume matches the
actual bottle filling and set the shutoff limit to a
reasonable value (e.g. 100 mL for 1 L bottles)
Probable cause Suggested actions
1 Upper pressure limit set too low.
Ensure the upper pressure limit is set to a value
suitable for the analysis.
2 Defective main board.
Please contact your Agilent service
representative.