Testing
1. Power On
" "
The LCD screen will appear for one second as unit performs
a quick diagnosis. A long tone indicates device is ready for
testing.
Press and hold button until a beep sounds.
Note: Unit will not function if residual air from previous
testing is present in cuff. The LCD will flash " "
until pressure is stabilized.
2.
Initial pressure is first pumped to 190mmHg. If the current
user's systolic blood pressure is over 190mmHg, the unit
will automatically re-inflate to the proper shelf.
Pressurization
Note: Pressurization will gradually subside and ultimately
stop when cuff is not properly applied to the wrist.
If this occurs, pressing the button to
turn the unit off.
" "
3. Testing
After cuff inflation, air will slowly subside as indicated by the
corresponding cuff pressure value. A flashing " " will
appear simultaneously on screen signaling heart beat detection.
Note: Remain relaxed during testing. Avoid speaking or
moving body parts.
4. Result Display
Three short beeps sound when testing is complete.
The screen will display measurements for systolic and
diastolic blood pressure. An indicator representing
the current measurement will appear.
Indicator
mmHg
M
D
mmHg mmHg
M
D
Unit O p er a t ion Unit O p er a t ion
2221
Memory Check
With power off, you may check past test results by using the
" M" button.
Upon activating test results, you can press the "M "button
to scroll through all test results stored in memory. The LCD will
display the last memory as NO reading.measurement
Power Off
The " " button can be pressed to turn off the unit
in any mode.
The unit can turn off the power itself in about 3 minutes with no
operation in any mode.
Safety Precaution: If pressure in cuff becomes too extreme
while testing, press the " " button to turn
power off.
The cuff pressure will rapidly dissipate once the unit is off.