5 Testing the Installation
To ensure that the powerchair meets a minimum level of safety, the following
procedure should be undertaken to ensure that the powerchair operates safely.
This procedure should be carried out in a spacious environment and with due
regard to any possible unexpected powerchair movement in the event of faulty
installation.
1. Turn the powerchair on by pressing the Power button on the DX Master
Remote.
2. Check that the green DX-ARC-SWB ‘Status’ light turns on and is
steady (i.e. not flashing).
3. Operate each seat switch in turn and check for the correct operation of
all seat functions.
4. Check that all other chair functions operate normally.
6 Specifications
6.1 Electrical Specifications
Min Nominal Max Units
General
Operating Voltage 18 24 32 Volts
Quiescent Current, Off 1 mA
Actuator Switch Inputs
Switch Closed (ON) Resistance 0 Short Circuit
100 Ohms
Switch Closed (ON) Current 2.5 10 mAmps
Switch Open (OFF) Resistance 5000 Open Circuit
Ohms
Auxiliary Power Supply
Output Voltage V
BATT
- 1 V
BATT
Volts
Output Current 1 Amps