PSI 5000 A Series
2.3.9 Initial commission
Fortherststart-upafterpurchasingandinstallingthedevice,thefollowingprocedureshavetobeexecuted:
• ConrmthattheconnectioncablesforDCtobeusedareofasatisfactorycrosssection
• Check that the default settings for set values, safety and monitoring functions and communication are suitable
for your application and change them where necessary, as described in the manual
• In case of remote control via PC, read the additional documentation for interfaces and software
2.3.10 Initial network setup
The device is delivered with default network parameters. The Ethernet/LAN port is immediately ready for use after
the initial commission. Default parameters:
IP: 192.168.0.2
Subnet mask: 255.255.255.0
Gateway: 192.168.0.1
Port: 5025
DHCP: off
For wiring, i.e. the hardware connection to a network, contact and ask your IT manager or any similar responsible
person. Network cable of common type (CAT5 or better) can be used.
In order to set up the network parameter to your needs, you have two options: the device’s website or the software
„EA Power Control“, which is available since April 2015 as download or included with your device.
Forthecongurationviathedevice’swebsiteorEA Power Control, you need the device to be connected to a
network or directly to a PC which can access the default IP 192.168.0.2.
How to do the network setup on the device website
1. Open the device website in a browser by entering the default IP (http://192.168.0.2) or the default host name
(http://Client, only possible if there is a running DNS in the network) into the URL box.
2. Afterthewebsitehasbeencompletelyloaded,checkthestatuselditemAccess” to show the status “free”.
In case it shows different, the device is either already in remote control (rem) or blocked from remote control
(local). If it shows “local”,rstremovetheblock.Refertosection„3.5.2. Control locations“ to do that.
3. If it says “rem” in the “Access” item, skip to step 4. Else enter command syst:lock on (attention! space
before on) into the SCPI command box and send with return key. Check if item “Access”instatuseld
changes to “rem-eth” (means: remote over Ethernet).
4. Switch to page CONFIGURATION (upper left corner) and set up the network parameters as well as the port
here resp. activate DHCP and submit the change with SUBMIT button.
5. Wait a few seconds before testing the new IP by entering it in the browser’s URL box. Opening the website
again by using the host name is only possible after the device has restarted, because only then the new IP
is reported to the DNS.
2.3.11 Commission after a rmware update or a long period of non-use
Incaseofarmwareupdate,returnoftheequipmentfollowingrepairoralocationorcongurationchange,similar
measures should be taken to those of initial start up. Refer to „2.3.9. Initial commission“.
Only after successful checking of the device as listed may it be operated as usual.