In case the DC output is already switched on, toggling the pin will switch the DC output off, similar to what it does
in analog remote control:
DC-
output
Pin
„REM-SB“
+
Parameter
„REM-SB“
Behaviour
is on
HIGH
+
normal
DC output remains on, nothing is locked. It can be switched on
or off by pushbutton or digital command.
LOW
+
inverted
HIGH
+
inverted
DC output will be switched off and locked. Later it can be switched
on again by toggling the pin. During lock, pushbutton or digital
command can delete the request to switch on by pin.
LOW
+
normal
b) Remote control of current and power
Requires remote control to be activated (Pin “Remote” = LOW)
The set values PSEL and CSEL are generated from, for example, the
reference voltage VREF using potentiometers for each. Hence the
power supply can selectively work in current limiting or power limit-
ingmode.Accordingtothespecicationofmax.5mAfortheVREF
output,potentiometersofatleast10kΩmustbeused.
The voltage set value VSEL is to VREF and will thus be permanently
100%.
If the control voltage is fed in from an external source it is necessary to
consider the input voltage ranges for set values (0...5 V oder 0...10 V).
Use of the input voltage range 0...5 V for 0...100%
set value halves the effective resolution.
Example with external
voltage source
Example with
potentiometers
c) Reading actual values
The AI provides the DC output values as current and voltage monitor. These can be
read using a standard multimeter or similar.