F Series Fetal & Maternal Monitor User Manual Maternal Monitoring
- 102 -
9.4.9 Changing SYS Alarm Limits
You can change the SYS alarm limits.
1 Select the setup key
on the main interface.
2 Select Alarm. On the displayed Password box, enter 9999, then select Enter.
3 Select SYS.
4 Select a value from 40 ~ 269 for Low Alarm Limit. (The step is 1, and the default value is
90 mmHg.)
5 Select a value from 41 ~ 270 for High Alarm Limit. (The step is 1, and the default value is
160 mmHg.)
6 Select OK.
9.4.10 Changing DIA Alarm Limits
You can change the DIA alarm limits.
1 Select the setup key
on the main interface.
2 Select Alarm. On the displayed Password box, enter 9999, then select Enter.
3 Select DIA.
4 Select a value from 10 ~ 214 for Low Alarm Limit. (The step is 1, and the default value is
50 mmHg.)
5 Select a value from 11 ~ 215 for High Alarm Limit. (The step is 1, and the default value is
90 mmHg.)
6 Select OK.
9.4.11 *Choosing NIBP Printing Mode
When the recorder is printing real-time fetal traces, the NIBP result is also recorded on the paper
whenever NIBP measurement is performed. After the paper stops advancing, you can choose to
keep recording NIBP results on the paper.
To enable or disable NIBP printing after paper advancing stops,
1 Select the setup key
on the main interface.
2 Select Recorder > NIBP.
3 Select ON or OFF (default).
4 Select OK.