User's Manual

F9, F9 Express Fetal & Maternal Monitor User Manual FTS-3 Fetal Telemetry System
- 122 -
11.2.5 Connect to the Base Station
1. Power on the base station.
2. Connect one end of the signal cable to the base station and the other end to the monitor
input socket.
3. Put the transducer back into the docking slot. The system can support 2 US transducers
and 1 TOCO transducer at most. Please do not exceed the maximum number of the
transducer.
NOTE:
If the system is provided with transducer protection cover, please do not take up the
cover during monitoring.
11.2.6 Configure the Monitor
1. Charge the transducer battery.
2. Power on the monitor.
3. Achieve the fetal heart signal.
Take the transducer up and keep the transducer at a distance of over 30cm from the base
station. The wireless connection indicator is on, and it indicates the transducer is taken
out. If you want to power off the transducer, put it back in the docking slot. If the
transducer connects to the base station successfully, the wireless connection indicator is
always on and do not put back the inactivated transducer in the docking slot.
4. Place the transducer on the patient.
NOTE:
1. Detailed operations please refer to 7.2.2 FHR Monitoring Procedure.
2. If the working status indicator is on, please do not put the uncharged transducer in
the docking slot.
3. The transducer has been taken first displays US1 on the screen, and that taken later
displays US2. Please do not take two US transducers simultaneously and wait at 2
seconds to take the other one.
11.2.7 Adjusting the Working Channel
If the fetal heart sound is with interference or it cannot be played smoothly, the working channel
is probably interfered. Put all the transducers back in the docking slots and press the adjustment
button
in back of the base station. The channel range is 1-32.
Restart the system when it enters the charging interface.