User's Manual

F9, F9 Express Fetal & Maternal Monitor User Manual FTS-3 Fetal Telemetry System
- 125 -
water and coupling gel during charging.
When you charge the battery, the screen will display as follows:
x Full charging icon: fully charged.
x Increasing charging icon: charging
x No charging icon: the transducer place in the docking slot incorrectly.
x If the screen displays ERROR, it indicates that the transducer is not connected well or
you place the transducer from the other system by mistake.
It takes about 3.5 hours to charge the battery. It is recommended to place the transducer in the
docking slot when the transducer is not used for a long time.
Install the transducer in the base station and the transducer icon will display on the screen.
At the end of their life hand the batteries over to the applicable collection points for the recycling
of waste batteries. If the battery charging time decreases sharply, the battery is considered as
obsolete battery. Please use the battery provided by the manufacturer and disposes the battery
according to the local regulations.
11.4.3 Basic Function Test
Please test the system after each service.
1. Power on the base station and connect it to the fetal monitor.
2. Charge the transducer.
3. Power on the monitor.
4. Take up the US transducer and test the following function:
x The US transducer screen displays the standard start interface.
x The US transducer indicator is green.
x The fetal monitor screen displays US.
5. Simulate the audio frequency signal:
x The fetal monitor displays FHR.
6. Take up the TOCO transducer and test the following function:
x The TOCO transducer screen displays the standard start interface.
x The TOCO transducer indicator is green.
x The monitor screen displays TOCO.