CAUTION
1 Although the monitor and the base station are chemically resistant to most common
hospital cleaners and non-caustic detergents, different cleaners are not
recommended and may stain the monitor.
2 Many cleansers must be diluted before use. Follow the manufacturer’s directions
carefully to avoid damaging the monitor
and the base station.
3 Do not use strong solvent, for example, acetone.
4 Never use an abrasive such as steel wool or metal polish.
5 Do not allow any liquid to enter the product, and do not immerse any part of the
monitor into any liquid.
6 Avoid pouring liquids on the monitor while cleaning.
7 Do not allow any remaining solution on the surface of the monitor.
NOTE:
1 The monitor surface can be cleaned with hospital-grade ethanol and dried in air or
with crisp and clean cloth.
2 The manufacturer has no responsibility for the effectiveness of controlling infectious
disease using these chemical agents. Please contact infectious disease experts in
your hospital for details.
13.2.2 Cleaning of Accessories
(1) Cleaning of Transducers
To clean the transducers and leads, follow these steps:
1) Wipe them with a soft cloth dampened in cleaning solution;
2) Clean them with a soft cloth dampened in water;
3) Air-dry them or wipe the remaining moisture with a soft dry cloth.
The recommended cleansers for accessories are listed below:
Accessory Cleansers
Ultrasound Transducer
TOCO Transducer
(Including the wireless)
Mild near neutral detergent Ethanol 75% Isopropanol
70%
DECG Leads Mild near neutral detergent Ethanol 75% Isopropanol70%
IUP Cable Mild near neutral detergent Ethanol 75% Isopropanol 70%
ECG Leads Mild near neutral detergent Ethanol 75% Isopropanol70%
SpO
2
Sensor Mild near neutral detergent Ethanol 75% Isopropanol70%