F Series Fetal & Maternal Monitor User Manual Maintenance and Cleaning
- 135 -
CAUTION
1 Do not squeeze the rubber tube on the cuff.
2 Do not dry-clean the cuff.
3 Only clean the outer surface of the connectors, make sure no liquid goes into the
connector.
4 When the reusable cuff is not connected with the monitor, or being cleaned, always
place the cover on the rubber tube to avoid liquid permeation.
13.3 Disinfecting
To disinfect the transducers and leads, follow these steps:
1) Clean the accessories.
2) Wipe them with a soft cloth dampened in the recommended disinfectant.
3) Wipe them clean with a soft cloth dampened in water.
4) Air-dry them or wipe the remaining moisture with a soft dry cloth.
The table below lists the allowed disinfectant bases:
Type Recommended
Fetal/Maternal Monitor
Ethanol 75%
Isopropanol 70%
Ethanol 75%
Isopropanol 70%
Base Station
US and TOCO Transducers
(the wired and wireless)
Remote Event Marker
DECG Cable
IUP Cable
ECG Leads
SpO
2
Transducer
TEMP Transducer
NIBP Cuff
NIBP Cuff Extension Tube