User's Manual

F Series Fetal & Maternal Monitor User Manual Product Specifications
- 148 -
Disinfection/Sterilizing
Method
Refer to this user manual for details
EMC
CISPR11 Group 1 Class A
Ground Leakage Current
(Limit):
N.C. S.F.C.
500μA 1000μA
Enclosure Leakage Current
(Limit)
N.C. S.F.C.
100μA 500μA
Patient Leakage Current
(Limit):
N.C. S.F.C.
d.c. 10μA 50μA
a.c. 100μA 500μA
Patient Auxiliary Current
(Limit):
N.C. S.F.C.
d.c. 10μA 50μA
a.c. 100μA 500μA
Base Station
Weight:
1.8 kg
Size:
310mm x 235mm x81mm
US Transducer
Weight:
About 150 g
Size:
Ø81 mm × 35 mm
TOCO Transducer
Weight:
About 150 g
Size:
Ø81 mm × 35 mm
B1.3 Performance Specifications
Ultrasound
Technique: Ultrasound Pulse Doppler with autocorrelation
Pulse Repetition Rate:
2 KHz
Pulse Duration:
92 μs
Ultrasound Frequency:
(1+
10%) MHz