User's Manual

F Series Fetal & Maternal Monitor User Manual Abbreviation
- 169 -
Operating Mode: PW mode
Working Frequency: 1.0MHz
Acoustic Output MI
I
spta.3
(mW/cm
2
)
I
sppa.3
(W/cm
2
)
Global Maximum Value 0.01087 0.6186 0.003412
Associated Acoustic
Parameter
P
r.3
(MPa) 0.01087
W
0
(mW) 3.496 3.496
f
c
(MHz) 1.00 1.00 1.00
Z
sp
(cm) 11.80 11.80 11.80
Beam
dimensions
X
-6
(cm)
0.5663 0.5663
Y
-6
(cm) 0.5437 0.5437
PD (usec) 90.65 90.65
PRF (Hz) 2000 2000
EDB
A
z
(cm)
3.46
E
le
(cm)
3.46
Operating Control
Conditions
Fixed