F Series Fetal & Maternal Monitor User Manual EMC Information
- 176 -
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(> 95% dip in U
T
)
for 5 sec
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(> 95% dip in U
T
)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the F Series Fetal & maternal
Monitor requires continued
operation during power mains
interruptions, it is recommended
that the F Series Fetal &
maternal Monitor be powered
from an uninterruptible power
supply or a battery.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.