Note the following:
These remote functions will apply only if
the remote controls are enabled through
this I/O Module.
Lines 4 and 5 control only the desired
range of the span function as described
in Line 2.
If the digital lines described above are
not connected, the Line 2-related span
function will only select Range 1; it will
not calibrate any other range. However,
the Line 3 function will span or check
ranges as setup in the module's pro-
gramming.
After calibration, the Analyzer module
will be returned to the range it was op-
erating on previous to calibration.
ITEM FUNCTION NOTES
Output Analog, 0 to 5 V, 0/4 to 20 mA
Support independent range se-
lection from that of the Analyzer
(no autorange)
Alarm relay 1 Calibration in progress SPDT
Alarm relay 2 Calibration gas / sample gas SPDT
Alarm relay 3 Zero gas valve SPDT
Alarm relay 4 Span gas valve 1 SPST
Alarm relay 5 Span gas valve 4 SPST
Alarm relay 6 Span gas valve 3 SPST
Alarm relay 7 Span gas valve 4 SPST
Input line 1 Perform a zero cycle
Input line 2 Perform a span cycle
Input line 3 Perform a complete calibration cycle
Input line 4 Range selection
Input line 5 Range selection
Input line 6 Calibration / Check selection
Table 11-6. Available Functions – Single Analyzer Autocalibration I/O Module V2.2