Rosemount Analytical Inc. A Division of Emerson Process Management Appendices 11-35
Model NGA 2000 Platform
Troubleshooting
No Output From Module
Verify that the module is marked as present in
the Local I/O setup menu. Verify that the right
number of analog output sub-modules are
present. Verify that the output is assigned to a
real signal from a real Analyzer.
No Digital Output
Verify that the serial output sub-module is
present. Verify that the module itself is
marked as present in the Local I/O module
setup menu. Verify that the communication
settings are correct in the Serial interface
setup menu. Verify that AK or Modbus is se-
lected in this menu. Make sure that you have
a standard serial cable to connect to this
module. Verify that the PC (or whatever de-
vice is used) serial settings match those of the
Local I/O module.
PIN SIGNAL NAME
1 Voltage output 1
2 Current output 1
3 Voltage output 2
4 Current output 2
5 Ground for Voltage output 1 & 2
6 Ground for Current output 1 & 2
7 Voltage output 3
8 Current output 3
9 Voltage output 4
10 Current output 4
11 Ground for Voltage output 3 & 4
12 Ground for Current output 3 & 4
13 Shield
14 Voltage output 5
15 Current output 5
16 Voltage output 6
17 Current output 6
18 Ground for Voltage output 5 & 6
19 Ground for Current output 5 & 6
20 Voltage output 7
21 Current output 7
22 Voltage output 8
23 Current output 8
24 Ground for Voltage output 7 & 8
25 Ground for Current output 7 & 8
Table 11-9. Local I/O 25-Pin Sub-D Connector Pin Identification