Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9
Model NGA 2000 Platform
CLD 7.50 ppm
HOME
ESCAPE MORE INFO
-- Current measurement
p
arameters --
Response time: 3.0 s
Bypass flow: 1300 ml/min.
Capillary pressure: 4.0 hPa
Detector pressure: 38.8 c
CLD 7.50 ppm
HOME
ESCAPE MORE INFO
-- Current measurement
p
arameters --
Analyzer gas measured: NO
Measurement range number: 1
Range change control: Local
Linearization mode: Disabled
Analyzer operational state: STANDBY
Analyzer alarm state: NORMAL
Alarms reported: FAILURE
Current total variable updates per second: 24
CLD
Sample flow: 1300 ml/min 200 1500
Sample press: 4.0 hPa 50.0 490.0
Ozonator: OFF – PRESS. SW.
Converter temp: 30.9 C 150.0 500.0
Dis
p
la
y
Status Main… Channel BasicCal
0.00 Ran
g
e: 1 10.00
7.50
ppm NO
It is normal for a Platform to run out of
memory if it is given too many Analyzer
modules to deal with, or Analyzer mod-
ules are replaced too often without reset-
ting it. V2.2 Platforms can deal with about
6 Analyzer modules, and V2.3 or later
with about 10 (when used with the in-
creased RAM Platform Controller board).
The symptom of this is that the Platform
hangs up while starting up, or continu-
ously restarts itself. In this case, you need
a hardware upgrade, and should call
Rosemount Service.
3-6 BASIC CONTROLS, SETUP AND STATUS
The following sections describe the basic
control of the NGA 2000 Platform and any in-
stalled Analyzers. Examples of stepping
through the menus are shown so that the user
can become familiar with the operation,
keeping in mind that displays and menu
choices may be different depending on the
configuration and any customization of the
menus.
NOTE
See individual Analyzer manuals for
more detail and module specifics.
Figure 3-10. Analyzer Channel Status Display
If necessary, from the Main menu press
MEASURE (F1).
Press STATUS (F2) to change to the
Analyzer Channel Status menu.
Or, from the Main menu press STATUS (F2)
to change to the Analyzer Status menu.
From the Current measurement parameters
menu an additional sub menu is available
with additional parameters.
Press MORE (F3) to access the next menu.
Press ESCAPE (F2) or MORE (F4) to return
to the first menu or HOME (F1) to return to
the Main menu display.
Different Analyzer modules will present
different Status menu displays.