Instruction manual

Instruction Manual
760006-A
September 2001
4-4 System SIO Module Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 Platform
a. Analog Output
Selecting the analog output
number
Figure 4-3. SIO Analog Output Setup - Selecting The Analog Output Number
Selecting the Analyzer module
Figure 4-4. SIO Analog Output Setup - Selecting Analyzer module
CLD 7.50 ppm
Measure
Back…
-- Anal
y
zer Modules --
CLD : 1.0
MLT/CH1
MLT/CH2
MLT/CH3
>>><<<
CLD 7.50 ppm
Measure
Back…
-- Analo
g
Output Setup --
Output number: 1
Choose signal source mdoule…
Choose signal…
Signal value for 0% output: 0.00
Signal value for 100% output: 100.00
Output current: 0…20 mA
Hold output during calibration: No
Signal name: Sample flow
Current signal value: 8.60
Source module: CLD
More…
Select the tag of the desired reference
channel and press Enter (). The display will
return to the Analog Output Setup menu and
the selected reference channel will be
displayed in the Source Module… line.
(The available selections may be different
depending on the installed and selected
module.)
Select Choose signal source module… and
press the Enter () key.
Choose the desired analog output (1-8) to set
the parameters. The number of outputs
depends on the Analyzer configuration as 2,
4, 6, or 8.
CLD 7.50 ppm
Measure
Back…
-- Analo
g
Output Setup --
Output number: 1
Choose signal source mdoule…
Choose signal…
Signal value for 0% output: 0.00
Signal value for 100% output: 100.00
Output current: 0…20 mA
Hold output during calibration: No
Signal name: Sample flow
Current signal value: 8.60
Source module: CLD
More…