4-8 System SIO Module Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 Platform
Output Signal If Assigned Module Fails
From the Analog Output Setup menu,
press More… (F5), the display is now
Output Signal if Assigned Module Fails
and Fine Adjustment.
Output value on Analyzer failure: Choose
the desired signal level to cause a failure
condition. The Choices are:
Actual
BeginOfRange
EndOfRange
BeginOfRange-10%
BeginOfRange+10%
Fine Adjustment
Choose the output number (1-8) for set-
ting the fine adjustment.
Figure 4-10. SIO Analog Output Setup – Output Signal If Assigned Module Fails/Fine Adjustments
In the Analog Output Setup menu, pressing
the More… (F5) key changes to the Output
Signal if Assigned Module Fails and Fine
Adjustment menus (both displayed on same
screen)
CLD 7.50 ppm
Measure
Back…
-- Analo
g
Output Setup --
Output number: 1
Choose signal source mdoule…
Choose signal…
Signal value for 0% output: 400.00
Signal value for 100% output: 700.00
Output current: 0…20 mA
Hold output during calibration: No
Signal name: Sample flow
Current signal value: 8.60
Source module: CLD: 1.0
More…
CLD 7.50 ppm
Measure
Back…
-- Output Signal if Assigned Module Fails --
Output(s) value on analyzer failure: BeginOfRange – 10%
More…
-- Fine Ad
j
ustments --
Output number: 1
Operation mode: Normal
Fine adjustment for 0% output: 4096
Fine adjustment for 100% output: 819