Instruction manual

Instruction Manual
760006-A
September 2001
5-4 System DIO Module Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 Platform
Figure 5-2. System DIO Module – Configure Inputs (continued next page)
Main Menu
Analyzer and I/O, expert controls and setup…
System & network I/O module controls…
In the DIO Module Inputs menu, use the
cursor keys to go to Input number:, press
Enter (). Select the desired input number
in the range of 1-8.
Using the cursor keys, go to Choose
module… menu line, press Enter ().
In the System & Network I/O Module
Controls menu, use the cursor keys to
go to System DIO module… menu line,
press Enter ().
(If there is no corresponding DIO module
installed, a corresponding message will be
displayed instead of a menu.)
CLD 7.50 ppm
Measure
<<< Back… >>>
-- S
y
stem & Network I/O Module Controls --
System SIO module…
System DIO module…
CLD 7.50 ppm
Measure
Next… Back… Ackn!
-- DIO Module Outputs --
Inputs…
Output number: 1
Choose module…
Choose signal…
Invert signal: No
Module status: Normal
Slot ID: 1
Signal name: ???
Signal level 000.0
Signal comes from: CLD
CLD
Measure
Back…
-- DIO Module Inputs --
7.50 ppm
Input number: 1
Choose module…
Choose function…
Sort ID: 1
Signal name: ???
Signal level: 000.0
Signal comes from: CLD
Next…
CLD 7.50 ppm
Measure
<<< Back… >>>
-- Choose Source Module --
Control module 0.0
CLD 1.0
In the DIO Module Outputs menu, use the
cursor keys to go to Inputs… menu line,
press Enter ().
In the Choose Source Module menu, select
the desired source module for the digital
input number (1-8) being configured.
(The list of modules will depend on the
installed modules.)