5-4 System DIO Module Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 Platform
Figure 5-2. System DIO Module – Configure Inputs (continued next page)
Main Menu
↓
Analyzer and I/O, expert controls and setup…
↓
System & network I/O module controls…
In the DIO Module Inputs menu, use the ↑ ↓
cursor keys to go to Input number:, press
Enter (↵). Select the desired input number
in the range of 1-8.
Using the ↑ ↓ cursor keys, go to Choose
module… menu line, press Enter (↵).
In the System & Network I/O Module
Controls menu, use the ↑ ↓ cursor keys to
go to System DIO module… menu line,
press Enter (↵).
(If there is no corresponding DIO module
installed, a corresponding message will be
displayed instead of a menu.)
CLD 7.50 ppm
Measure
<<< Back… >>>
-- S
y
stem & Network I/O Module Controls --
System SIO module…
System DIO module…
CLD 7.50 ppm
Measure
Next… Back… Ackn!
-- DIO Module Outputs --
Inputs…
Output number: 1
Choose module…
Choose signal…
Invert signal: No
Module status: Normal
Slot ID: 1
Signal name: ???
Signal level 000.0
Signal comes from: CLD
CLD
Measure
Back…
-- DIO Module Inputs --
7.50 ppm
Input number: 1
Choose module…
Choose function…
Sort ID: 1
Signal name: ???
Signal level: 000.0
Signal comes from: CLD
Next…
CLD 7.50 ppm
Measure
<<< Back… >>>
-- Choose Source Module --
Control module 0.0
CLD 1.0
In the DIO Module Outputs menu, use the ↑
↓ cursor keys to go to Inputs… menu line,
press Enter (↵).
In the Choose Source Module menu, select
the desired source module for the digital
input number (1-8) being configured.
(The list of modules will depend on the
installed modules.)