Instruction manual

Instruction Manual
760006-A
September 2001
Rosemount Analytical Inc. A Division of Emerson Process Management System DIO Module 5-5
Model NGA 2000 Platform
Figure 5-3. System DIO Module – Configure Inputs (continued from previous page)
CLD 7.50 ppm
Measure
<<< Back… >>>
-- Functions --
AM: Zero-Cal
AM: Span-Cal
AM: Range-1
AM: Range-1
AM: Range-1
AM: Range-1
SYS: Zero-Cal
SYS: Zero/Span-Cal
CLD
Measure
Back…
-- DIO Module Inputs --
7.50 ppm
Input number: 1
Choose module…
Choose function…
Sort ID: 1
Signal name: ???
Signal level: 000.0
Signal comes from: CLD
Next…
Back in the DIO Module Inputs menu, use
the cursor keys to go to Choose
Function… menu line, press (Enter).
In the Functions menu, select the desired
signal for the digital output number (1-24)
being configured. If available, press
(Enter) for additional signals.
(The list of signals will depend on the
Analyzer module.)