Instruction manual
Instruction Manual
760006-A
September 2001
Rosemount Analytical Inc. A Division of Emerson Process Management System Configuration and Diagnostics 6-3
Model NGA 2000 Platform
6-2 AUTOCALIBRATION INTERVAL OPERA-
TION
In the example shown in Figure 6-2, the sys-
tem time is 17:00:00 May 29, 2001.
Therefore, the Zero Cal is set to start at hour
1 with an interval of 24 hours will next occur at
1 AM on the following day.
The Zero & Span Cal is set to start at hour 7
with an interval of 48 hours will next occur at 7
AM on the second day.
The Program Cal is set to start at hour 20 with
an interval of 24 hours will next occur at 8 PM
on the following day.
Start of interval time: is the time of day the
event is to occur in the range of 0 to 23.
Interval time: is the desired interval between
calibrations in the range of 1 to 10,000 hours
(up to 416 + days).
If the start of interval time is prior to the cur-
rent system time, the event will first occur at
the nearest start time plus the interval time.
Figure 6-2. Autocalibration Interval Operation Setup
From the System Calibration menu use the
↑ ↓ keys to go to the Times for interval
operation… menu line, press ↵ (Enter).
CLD 7.50 ppm
Measure
Back… Result…
-- System Calibration --
Calibration and test procedures…
Calibration sequence programming…
Times for interval operation…
Analyzer module setup…
CLD 7.50 ppm
-- Times for Interval O
p
eration --
Zero Cal is: Enabled
Start of interval time: 1 h
Interval time: 24 h
Zero & Span Cal is: Enabled
Start of interval time: 7 h
Interval time: 48 h
Program Cal is: Enabled
Start of interval time: 20 h
Interval time: 24 h
Next calibration events…
Measure
Back…
CLD 7.50 ppm
-- Next Calibration Events --
Zero Cal: 01:00:00 May 30, 2001
Zero & Span Cal: 07:00:00 May 31, 2001
Program Cal: 20:00:00 May 30, 2001
Measure
Back…
Main Menu
↓
System configuration and diagnostics…
↓
Sy
stem calibration…
In the Times for Interval Operation menu,
enable or disable the three available
calibration events as desired.
To verify when the enabled events will occur,
use the ↑ ↓ keys to move to the Next
calibration events… menu line, press ↵
(Enter).