6-6 System Configuration and Diagnostics Rosemount Analytical Inc. A Division of Emerson Process Management
Model NGA 2000 Platform
6-5 CALIBRATION RESULTS
Displays the results of the last calibration.
Figure 6-5. Display Calibration Results
CLD 7.50 ppm
Measure
Back… Result…
-- System Calibration --
Calibration and test procedures…
Calibration sequence programming…
Times for interval operation…
Analyzer module setup…
CLD 7.50 ppm
-- Calibration Results --
Choose analyzer module…
Analyzer module: CLD
Enabled for system calibration: Yes
Result of last run: OK
Successful zero+span calibrated ranges: None
Zero calibration failed on any analyzer module: Yes
Span calibration failed on any analyzer module: No
Measure
Back…
CLD 7.50 ppm
-- Calibration & Test Procedures --
Start Zeroing all ranges!
Start Zeroing and spanning all ranges!
Start calibration program!
Cancel calibration!
Calibration procedures in test mode: No
Test procedures…
Calibration type: NoOp
Program step: 1
Calibration time: 0 s
Previous calibration time: 120 s
Measure
Back…
Result…
From the System Calibration menu use the
keys to move to Calibration & test
procedures… menu line, press (Enter).
Press Result… (F5) key to display the
results of the last calibration for any bound
Analyzer module.
Choose the desired Analyzer module for
which to display the last calibration results.
The results are displayed on the four lines
below the line.