Instruction manual

Instruction Manual
760006-A
September 2001
Rosemount Analytical Inc. A Division of Emerson Process Management System Configuration and Diagnostics 6-13
Model NGA 2000 Platform
Bind Modules
Figure 6-10. Network Module Management – Bind Modules
In the Network Module Management menu,
use the keys to go to Bind modules…
line and press the (Enter) key.
The Module Binding screen displays the list
of one or more installed Analyzers. The list
may show I/O modules already bound to
Analyzers.
To add an I/O module to an Analyzer mod-
ule, highlight the desired Analyzer module
and press the Add! (F2) key to display the
Select I/O Modules screen which lists avail-
able I/O modules.
Use the <<< and >>> keys to scroll through
additional screens.
Highlight a listed I/O module, press the
(Enter) key, the display will return to the
Module Binding screen.
The selected I/O module will be indented
below the Analyzer module.
CLD 7.50 ppm
-- Network Module Mana
g
ement --
Measure
Back…
List of active modules…
Memory usage
Automatic bind of I/O-modules: Yes
Bind modules…
Erase inactive modules
Replace modules…
Note: Re-initializing will destroy all the binds.
Re-initialize network!
CLD 7.50 ppm
-- Module Bindin
g
--
Abort
Delete!
CLD1 CLD: 1.0
Bind!Add!
CLD 7.50 ppm
-- Select I/O Modules --
Measure
Back…
Select the modules you wish to bind to the current
analyzer.
Analog I/O
Autocal I/O
>>><<<
CLD 7.50 ppm
-- Module Bindin
g
--
Abort
Delete!
CLD1 CLD: 1.0
Analog I/O
Bind!Add!
Main Menu
System configuration and diagnostics…
Network module management…