Owner's manual

32
Note that the rcf (g-force) indicated when the display is switched is standardized to 15 ml Falcon
®
tubes in the
rotor A-4-38. At 4,400 1/min you can achieve the following maximum rcf with the various adapters.
To ascertain the maximum rcf for a specific adapter you can apply the following formula as per DIN 58 970:
rcf = 1,118 · 10
–5
· n
2
· r
max
n: rotational speed in 1/min
r
max
: max. centrifuging radius in cm
Example: The 85 ml adapter has a maximum radius of 13.5 cm. At 4,000 rpm a maximum rcf of 2,415 x g is reached.
time
The continuous run function is set by turning the TIME dial to either above
99 minutes or below 0.5 minutes. The timer displays "oo" to indicate that continuous
running is active. The time is counted upward in minutes.
stop
Ends continuous running
SPEED Press the dial to toggle the display between 1/min (rpm) and (rcf) and vice versa.
Rotor Adapter Max. centrifuging
radius (cm) r
max
Max. rcf (g-force),
rounded
A-4-38
Round bucket
1.5/2.0 ml 13.4 2,900
1.1 – 1.4 ml 13.2 2,850
2 – 7 ml 13.2 2,850
2.6 – 7 ml 13.2 2,850
4 – 10 ml 13.2 2,850
9 – 15 ml 13.0 2,800
15 ml Falcon
®
13.7 3,000
25 ml 13.5 2,900
50 ml Falcon
®
13.5 2,900
85 ml 13.5 2,900
A-4-38
Rectangular bucket
5 – 7 ml 12.7 2,750
9 ml 12.7 2,750
15 ml 12.7 2,750
20 ml 12.7 2,750
25 ml 12.7 2,750
A-8-17 15 ml Falcon
®
12.8 2,770
15 ml DIN-tube 12.8 2,770
F-35-30-17
Outer ring
15 ml 12.7 2,750
15 ml Falcon
®
12.7 2,750
F-35-30-17
Inner Ring
15 ml 10.7 2,300
15 ml Falcon
®
10.7 2,300
F-45-24-11 0.2 ml 6.6 1,430
0.4 ml 8.2 1,770
0.5/0.6 ml 7.4 1,600
1.5/2.0 ml 8.2 1,770
F-45-18-17-Cryo without adapter 9.1 1,970
with adapter 8.9 1,930
3.10 Continuous running
3.11 Rcf display and calculation
3 Operation
Operation