User manual

X K 3 1 9 0C 602
For instance, if the indicator measures a gross weight of 50.00 (kg), it will send data as
follows:
“G 50.00”;
If the indicator measures a net weight of 0.040 (kg), it will send data as follows:
“N 0.040”.
2. Command response mode
When SET 1 parameter 6B is set as 0 or 2, the serial interface 2 works in command
response mode. If serial interface 2 works in continuous send mode, when the PC sends correct
command to the indicator in command response mode, the indicator will automatically switch to
command response mode. If the PC does not send read in EEPROM command, C602 will still
work in continuous send mode after next startup. With regard to response mode, the PC sends
“setting the communication mode into continuous send mode”, the indicator will switch to
continuous send mode.
Please refer to appendix E for the communication format of command response mode.
12 0x0A Linefeed sign