X K 3 1 9 0C 602
SET 1 general parameters
Table 4-3 Table for Setting of General Parameters
Param
eter
Indicator
display
Parameter explanation Remarks
1
[ABCDEF]
[COMM ]
Hardware selection
Acommunication (1)
(0: communication (1) OFF, 1
:
communication (1) ON)
Bcommunication (2)
(0: communication (2) OFF, 1
:
communication (2) ON)
Cscoreboard
(0: scoreboard 1 OFF: scoreboard ON )
Dprinter
(0: printer OFF, 1: printer ON)
E analog quantity (0: Analog output
OFF, 1: Analog output ON)
Fdisplay of analog quantity
(0: net weight, 1: gross weight)
Relevant parameters of zero
Azero-setting upon startup (0: off, 1: on)
Bmanual zero-setting range (0 ~ 5)
Czero-setting range upon startup (0 ~ 5)
B, C 0 1 2 3 4 5
Max%
0 2 4 10 20 100
Dzero tracking range ( 08 )
D
0 1 2 3 4 5 6 7 8
2
[ ABCD]
[0 Set ]
(e) 0 0.5 1 1.5 2 2.5 3 3.5 4
AD relevant parameters
Afiltering algorithm
(0: sliding window filter 1: first-
order
low-pass filter)
BAD sampling speed (0 ~ 4)
CAD filtering strength (0 ~ 4)
Dstable judgment (0 ~ 4)
Chara
cter
Meani
ng
0 1 2 3 4
3
[ ABCD]
[FLt ]
B
Rate
25 50 60 100 200
The larger the value of B
is,
the faster the AD
sampling
and the lower the stability
will be. T
he larger the value
of C
is, the greater the
stability
will be, but leading
to greater delay. T
he smaller
the value of D
is, the stricter
the stability
judgment will be,
that’s to say,
it is more