Judgment delay (0 ~ 25.5)
seconds:
U
nder external control mode,
when it is activated,
data
calculation is started
after
t0;
Under self-
control mode, data
calculation is started
after t0
when weight is out of
zero
zone.
Please refer to p
arameter 6
for the definition of โ€œ
zero
zoneโ€.
8
[ ***]
[t1 ]
T
ime for calculation of
average weight
(0 ~ 25.5) seconds:
After t0, the indicator
will
sum the weight values and
figure out the average
during time t1.
The obtained
value serves as
catchweighing basis.
A
ssume that the obtained
value is X.
9
[ ***]
[t2 ]
Calculation delay (0 ~
25.5)
seconds:
After t1
calculation is over,
the indicator
remains inactive
during t2 time, waiting
with
delay.
T
his setting can meet more
control site requirements.
10
[ ***]
[t3 ]
Time for sending signal
(0 ~ 25.5) seconds:
After t2, the indicator
sends
catchweighing signal with
length of t3
P
lease refer to the
explanation of parameter 2
~ parameter 5.
11
[ A]
[Print?]
Aโ€”
whether to print
parameter setting
(0: No: 1: Yes)
II. Calibration
Message: The calibration switch should be opened when the calibration is started,
otherwise the calibration status cannot be entered. Before exit by pressing the button [enter],
the calibration switch should be closed after completion of calibration; otherwise the
auxiliary display will show the message [CALEnd]; It can be quitted by pressing [exit] and