the calibration is valid.
Toggle the switch to ‘ON’, press the button Settingand press 5or 6repeatedly until
is shown. Press Input and the indicator will show , which means it is
the calibration status. Press Inputagain to start the calibration process. Please see the following
table for the calibration instruction and operation method (* is the original setting value).
Table 4-6 Calibration Setting
Step
Parameter
Display
Parameter Instruction Operation Instruction
1
[ *]
[dC ]
Decimal Digits0-3
Press Input
after
modifying the parameter
2
[ *]
[e ]
Division
1/2/5/10/20/50
Press Input
after
modifying the parameter
If the decimal digit is not
zero, 10, 20, 50 can
not be
selected.
3
[******]
[F ]
Max. Weighing
Press Input
after
modifying the parameter
4
[ A]
[CAL 0?]
ASave the original zero
point
0
Should reconfirm the
current zero point;
1
Skip the confirmation of
current zero point
Choose 0 to go to step 5;
Choose 1 to skip step
5 and
go to step 6.
5
[******]
[noLoAd]
The confirmation of zero
point
The upper part shows
AD
code of the indicator
Confirm that the current
weighing platform
has no
load and light is stable
ON,
then press Input
6
[******]
[AdLoAd]
Load weight
The upper
part shows AD
code of the indicator
Press Inputafter the load
is finished and the
light is
stable ON
7
[******]
[LoAd ]
Input the current load
weight
Press Input
after
changing
it to current
weight value
[ ]
[CALIb ]
[ 3]
[SEt ]