The calibration is done and
the calibration status is
quit
ted. Toggle the
calibration switch
to OFF
and pressInput
to go back
to normal weighing status.
III. Checking of calibration data
Prompt: The calibration switch should be opened when the calibration data is modified,
otherwise it cannot be saved.
Press Settingand 6 to show . Press Input, the indicator will enter
the data checking status, please see table 4-7.
Table 4-7 Calibration Data Checking
Parameter
Parameter
Display
Parameter Instruction
Operation Instruction
1
[ *]
[dC ]
Decimal Digits0-3
Press Input aft
er
modifying the
parameter
2
[ **]
[e ]
Division
1/2/5/10/20/50
Press Input
after
modifying the
parameter
If the decimal digit is
not zero, 10, 20, 50
can
not be selected.
3
[******]
[F ]
Max. Weighing
Press Input
after
modifying the
parameter
After modifying the
max. weighing
4
[******]
[0_Ad ]
Zero AD code
Press Input
after
modifying the
parameter
5
[******]
[bL ]
CAL Coefficient
Press Input
after
modifying the
parameter
6
[******]
[noLine]
Nonlinear Correction
ValueNote 1
(-1+1) Unit%Max
Press Input
after
modifying the
parameter
7
[ *]
[PZL ]
Tare
Cannot be modified
[ 4]
[SEt ]