User manual

X K 3 1 9 0C 602
8
[******]
[0Point]
Current Zero
(
Relative to calibration
zero)
Can
not be modified.
This parameter
reflects
the z
ero stability of
weighing load cell.
9
[ A]
[Print?]
A
Whether to print each
parameter
0No1Yes
Note 1: Nonlinear correction uses the parabola calculation method. The correction value is 1/2 max. weighing
position, that is, the correction amount at the top of the correction curve. Nonlinear correction value equals to the
negative value of the nonlinear tolerance of 1/2 max. weighing position. The unit of the nonlinear correction value
is 1% of the max. weighing value. For example, for the scale whose Max=10kg, if 5kg is added to it and the
weight shows 5006g, then the tolerance is +6g, that is , +0.06%Max and the nonlinear correction value is -0.06.