User manual

X K 3 1 9 0C 602
off quick feed. The timer will be T1 and go to program step 3.
3 The timing is over, go to program step 4 and the timer will be T0.
4 Start the slow feed; timing is over and weight inspection is started.
If the loaded value reaches to batching value minus slow feed lead,
turn off slow feed. Timer will be T2 and go to program step 5.
5 The timing is over. If the lead correction is allowed, modify the
slow feed lead and go to program step 6.
6 If the weight is qualified, the qualified signal will be output and it
will go to program step 9, the timer will be T5. If it is out of the
tolerance, output the out-of-tolerance signal. If the weight is less
than the batching value minus tolerance and the gradual feed is
allowed, go to program step 7 and the timer will be T3; If the
gradual feed is not allowed and the weight is more than the
batching value plus tolerance, out-of-tolerance treatment will stop
at this step and wait for qualified out-of-tolerance treatment. If the
out-of-tolerance treatment is not done, then go to program step 9,
the timer will be T5.
7 Start the slow feed, timing is over, the slow feed will be closed and
timer will be T4. Go to program step 8.
8 Treatment is same as program step 6
9 Timing is over, close the qualified or out-of-tolerance signal. If
material feeding action is started manually, then stop the action; If
it is under the status of automatic cycle, then go to program step
10.
10 Perform the automatic print and automatic storage operation
according to the setting and go to the program step 11.
11 Exit the status of negative scale, go to program step 12 and the
timer will be T7.
12 The timing is over. If the cycle time is not reached, go back to