User manual

X K 3 1 9 0C 602
Send by
PC
Write
calibration
parameter
02H Addr T NN
****** XH XL 03H
02H AT NN ******
03H
Refer to note 2, note 5.
T
Send by
instrument
Return to
calibration
parameter
02H
Addr T NN
****** XH XL 03H
02H AT NN ******
03H
Return to PC
command
Send by
PC
Write working
parameter
02H Addr U NN ******
XH XL 03H
02H AU NN ******
03H
Refer to Note 3Note 6
U
Send by
instrument
Return to working
parameter
02H Addr U NN ******
XH XL 03H
02H AU NN ******
03H
Return to PC
command
Send by
PC
Eliminate stored
data
02H Addr V XH XL
03H
02H AV17 03H
V
Send by
instrument
Eliminate stored
data
02H Addr V XH XL
03H
02H AV17 03H
Send by
PC
Read the date
02H Addr W X
H XL
03H
02H AW16 03H
W
Send by
instrument
Return to the
date
02H Addr W ******
XH XL 03H
02H AW05-07-
2214
03H
Send by
PC
Read the time
02H Addr X XH XL
03H
02H AX19 03H
X
Send by
instrument
Return to the
time
02H Addr X ****** XH
XL 03H
02H
AX17:09:27 33
03H
Send by
PC
Write the time
02H Addr Y ****** XH
XL 03H
02H AY ********
03H
Y
Send by
instrument
Return to the
time
02H Addr Y ****** XH
XL 03H
02H AY ********
03H
Send by
PC
Write the time
02H Addr Z ****** XH
XL 03H
02H AZ **
******
03H
Z
Send by
instrument
Return to the
time
02H Addr Z ****** XH
XL 03H
02H AZ ********
03H
Send by
PC
Select remote
control
02H Addr AA ******
XH XL 03H
02H AAA 071 03H
Note 7
AA
Send by
instrument
Select remote
control
02H Addr AA ******
XH XL 03H
02H AAA 071 03H
Send by
PC
Read IO state
02H Addr AB NN XH
XL 03H
02H AABI0 03H
Note 8
AB
Send by
instrument
Send IO state
02H Addr AB NN
****** XH XL 03H
02H AABI0********
03H