User manual

X K 3 1 9 0C 602
and data format are same as Note 8 of E-2.
Table E-2 Note 10: The meaning of NN parameter is shown in Table E-8.
Table E-8 Definition of Parameters
Parameter code Definition
TA~TH
Remaining time of time relay 0 ~ time relay 7; integer
of returned 3 bytes of C602 represents calculated value
of remaining time. Time unit 100ms, value range
0~255.
JA~JH
States of group 0 to group 7 intermediate relays, 8
intermediate relays for every group. Group 0 is time
relay. 8 digits of binary data represented by returned 3
digits of integer of C602 indicate the state of 8 relays.
PA~PZ
26 memories, including integers ranged
-8388608~8388607; long integer of returned 8 bytes of
C602
Note 11 of Table E-2: this command is only effective in remote control state. Parameter definition is shown
in Table E-7. Time relay state can not be written.
Table E-2 Note 12: returned character string is the content of corresponding relay.