User`s guide

22 MM560A-MM570A-EU-EN
V1.1 8/12
DC Current
RANGE Accuracy Burden Voltage
500.00
A
0.15%+20d
0.15mV/
A
5000.0
A
0.1%+20d
0.15mV/
A
50.000mA 0.15%+20d 3.3mV/mA
500.00mA 0.1%+30d 3.3mV/mA
5.0000A 0.5%+20d 45mV/A
10.000A* 0.5%+20d 45mV/A
*10A continuous, >10A to 15A (to 20A for MM570A) for 30 second max with 5 minutes cool down
interval
AC & AC+DC Current
MM570A
MM560A
RANGE
Accuracy
Burden Voltage
50Hz -- 60Hz
500.00
A 0.15mV/
A
5000.0
A 0.15mV/
A
50.000mA 3.3mV/mA
500.00mA 3.3mV/mA
5.0000A 45mV/A
10.000A*
0.5%
+50d
1.0%
+40d
45mV/A
40Hz – 1kHz
500.00
A 0.15mV/
A
5000.0
A 0.15mV/
A
50.000mA 3.3mV/mA
500.00mA 3.3mV/mA
5.0000A 45mV/A
10.000A*
0.7%
+50d
1.0%
+40d
45mV/A
1kHz – 10kHz
500.00
A 0.15mV/
A
5000.0
A 0.15mV/
A
50.000mA 3.3mV/mA
500.00mA
2.0%
+50d
Unspec'd
3.3mV/mA
5.0000A
10.000A*
Unspec'd Unspec'd 45mV/A
*10A continuous, >10A to 15A (to 20A for MM570A) for 30 second
max with 5 minutes cool down interval
DC Loop Current: %4--20mA, 4mA = 0% (zero), 20mA = 100% (span),
Resolution: 0.01%, Accuracy:
25d
Peak mode: Accuracy: Specified accuracy 100 digits for
changes > 0.8ms in duration
T1-T2 Dual Temperature (MM570A only)
RANGE
Accuracy
-50.0
C to 1000.0
C 0.3%+1
C
-58.0
F to 1832.0
F 0.3%+2
F