Datasheet

Specifications common to both
Dimensions 17.5cm H, 7.8cm W, 3.8cm D, .35kg weight including batteries
Operating temperature with the battery 0ºC to 50ºC
Non-operating temperature -20ºC to 60ºC
Operating relative humidity (without condensation) 95% (10ºC to 35ºC) 75% (35ºC to 40ºC) uncontrolled ‹ 10ºC
Vibration Random, 5Hz to 500Hz, MIL-PRF-28800F CLASS 2
Shock 1-meter drop test
Altitude 9,842 feet/3,000 metres
EMC EN 61326-1:2004
Battery type 2 AA alkaline batteries (no battery charger)
Battery life More than 1,500 tests (typical)
Laser safety and classification Class 1 CDRH Complies to EN 60825-2
LCD type Backlit black and white (segments)
Index of refraction range 1.45 to 1.5 (factory default is 1.468)
Auto turnoff Automatically turns off after 5 minutes if no keys are pressed. Backlight turns off first.
Factory calibration interval None
Maximum number of incidents shown 9
Testing speed 6 seconds typical testing time
Connector Removable/cleanable SC adapter, UPC
Loss Threshold Setting 1.5dB default warning threshold (user-configurable from 0.5dB to 6.1dB in 0.1dB increments)
Distance accuracy
± (1m + 0.1% x length) for reflective incidents
4
± (3m + 0.1% x length) for non-reflective incidents
4
Bulkhead quality If no fiber is attached or if the connector is dirty, the troubleshooter displays 0m or 0ft.
Certifications and compliance
Conforms to relevant European Union directives
Conforms to relevant Australian standards
Listed by the Canadian Standards Association CSA C22.2 No. 61010.1.04
Conforms to FCC Rules, Part A, Class A
RoHs
1a. Detects the location of an incident that has a reflectance larger than -62dB. Detects incidents ›2m after the bulkhead connector when the bulkhead reflectance is ‹35dB. Detects incidents ›3m after an incident when the incident reflectance is ‹35dB.
1b. Finds and gives the location of an incident that has a reflectance larger than -55dB. Detects incidents ›1m after the bulkhead connector when the bulkhead reflectance is 3 m after an incident when the incident reflectance is.
2. Detects incidents ›10m after the bulkhead connector or any prior incident when the bulkhead reflectance is ‹-35dB and the reflectance of any prior incident is ‹-35dB. The maximum link loss prior to the incident is -7dB
3. With a backscatter coefficient of -63dB at 850nm using a calibrated -14dB reference.
4. ± user-configurable Index of Refraction (IOR) error ± the incident location error.
6 of 8