DSX-600 Series CableAnalyzer
Users Manual
108
The serial numbers of the main and remote units are on a sticker
on the back of the units.
Before You Send a Main Tester to a Service
Center
WCaution
If you send a main tester to a Service Center, make
sure you download any test results you must keep
to LinkWare Live, LinkWare PC, or to a USB flash
drive. Service procedures can possibly delete test
results, and Fluke Networks is not liable for lost
results.
Table 6. Possible Solutions for Unusual Behavior
The touchscreen or the keys do not respond.
Press and hold until the tester turns off. Then turn on the tester. If the
problem continues, install the latest version of software in the tester.
The tester will not turn on even though the battery is charged.
The safety switch or the fuse in the battery is possibly open. Wait a few
seconds, then try again. If the tester does not turn on, the battery is possibly
bad and must be replaced. The tester can operate on AC power if the
battery does not operate.
Test results appear to be incorrect.
The tester possibly has incorrect settings. Make sure you selected the correct
cable type and test limit.