To set the reference:
1
Clean the connectors on the meter, source, and a test
reference cord. Use fiber optic solvent and optical-
grade wipes or swabs to clean connectors as described
on page 15.
2
Turn on the meter and source and let them warm up
for 5 minutes. Allow additional time if the equipment
has been stored above or below ambient temperature.
3
Make the connections shown Figure 10.
4
If the source’s AUTO LED is not on, press A.
If you want to set the reference for both wavelengths,
pre
ss Z until the wavelength LEDs blink alternately.
Note
If you are not using a SimpliFiber Pro source, set the
source
to the desired wavelength and continuous
wave output.
5
On the meter, press M until POWER appears (see
Figure 8 on page 23).
Notes
You cannot set the reference if Min or Ma
x appear
on the display. See page 24.
Setting the reference from power mode lets you
see the
source’s power level before saving it as the
reference level. You may also set the reference
from loss mode.
6
Press T SET REF. The meter switches to loss mode, the
display shows 0.00 dB, OK appears briefly, and the new
reference value appears.
If the reference value is less than -60 dBm or if the
source i
s in ID or 2 kHz mode, the meter briefly shows
and O. Check the connections and the source’s
mode or try setting the reference again using another
tes
t reference cord.
WCaution
If you disconnect the source’s output after setting
the reference, you must set the reference again to
ensure valid measurements.