User Manual

SimpliFiber Pro Meter and Test Kits
Users Manual
48
2 Set the Reference
2-1
Set the reference as described on page 26.
2-2
Make sure that the reference value is good:
For 50/125 µm fiber, the reference value must be
better than -24.50 dBm (for example, -23.50 dBm is a
better value)
For 62.5/125 µm fiber, the reference value must be
better than -20.00 dBm (for example, -19.50 dBm is a
better value)
For 9/125 µm fiber, the reference value must be
better than -8.00 dBm (for example, -7.50 dBm is a
better value)
If the value is not good, do the reference procedure
again with a different test reference cord.
3 Test the Cord in Both Directions
3-1
Disconnect the test reference cord from the meter
(Figure A-3,
A
).
3-2
Connect the cord you want test (Figure A-3,
B
).
3-3
On the meter, press M until LOSS appears.
3-4
If the source’s AUTO LED is not on, press A.
3-5
If you want the source to automatically switch
wavelengths, press
Z until the wavelength LEDs
blink alternately. Or you may press Z to switch
wavelengths as needed.
Note
If you are not using a SimpliFiber Pro source, set
the source to the desired wavelength and to
continuous wave output. On the meter, press S V
to select the wavelength generated by the source.
3-6
Compare the loss measurement to these limits:
Multimode test reference cord: 0.10 dB or less
Singlemode test reference cord: 0.20 dB or less
If the connectors show no contamination or damage,
but the loss is higher than these limits, the cord does
not comply with ISO/IEC 14763-3. Do not use the cord
as a test reference cord.
3-7
Connect the cord in the opposite direction (Figure A-3,
C
), then repeat step 3-6.