Product Manual

Product Overview and Specifications
Specifications 1
1-11
Specifications
General Specifications
Fuse Protection for Resistance ...................... 0.44 A (44/100 A, 440 mA), 1000 V FAST Fuse, Fluke specified part
only
Power Supply
Battery power ...................................................... BP500 smart battery pack: double cell lithium-ion, 7.4 V, 3000 mAh
Battery life ........................................................... >8 hours in continuous full-load operation
Battery charging time .......................................... 4 hours
Power adapter output voltage ............................. Use only BC500 battery charger: 18 V, 840 mA
Line power .......................................................... 100 V ac to 240 V ac adapter with country specific plug
Frequency ........................................................... 50 Hz to 60 Hz
Temperature
Operating ............................................................ 0 °C to 40 °C
Storage ............................................................... -20 °C to 50 °C
Lithium-ion battery charging ............................... 0 °C to 40 °C
Relative Humidity (non-condensing, 10 °C)
Operating ............................................................ 80 % at 10 °C to 30 °C
75 % at 30 °C to 40 °C
Altitude
Operating ............................................................ 2,000 m
Storage ............................................................... 12,000 m
Temperature Coefficient ....................................... 0.1 x (specified accuracy) /°C (<18 °C or >28 °C)
Size ......................................................................... 58 x 103 x 220 (mm)
Weight .................................................................... 850 g
Memory
Data/Setup flash memory ................................... 4 MB
Real-Time Clock .................................................... Time and date stamp for measurement. The RTC works >50 days
without battery.
IP Rating ................................................................ IEC 60529: IP 40
Safety ..................................................................... IEC 61010-1, IEC 61010-2-030, IEC 61010-031, Pollution Degree 2
600 V CAT III; Derated to CAT II with CAT II probe cap installed
EMI, RFI, EMC, RF ................................................. IEC 61326-1, IEC 61326-2-2
Electromagnetic Compatibility ............................ Applies to use in Korea only. Class A Equipment (Industrial
Broadcasting & Communication Equipment)
[1]
[1] This product meets requirements for industrial (Class A) electromagnetic wave equipment and seller or user should take notice of it.
This equipment is intended for use in business environments and is not to be used in homes.