Product Manual

Use the Product
Make Measurements 3
3-11
Set Measurement Thresholds
The Product lets you define upper and lower measurement thresholds or tolerance
ranges. These defined threshold values are then compared to the measured values to
automatically identify and prompt the user with a PASS, FAIL or WARN indicator of
battery out of tolerance conditions.
The threshold function is disabled by default. You can set up to 10 set of thresholds and
select one threshold as needed.
To set and select measurement thresholds:
1. On the measurement screen, push the Threshold softkey to open the Select
Threshold menu.
2. Use and to select one threshold set out of ten.
hpo033.jpg
3. Use and L to highlight the value to be edited among Voltage lower, Reference,
Warning, Fail, Low limit, and Notes.
4. Edit the selected field.
a. Use the and + softkeys to change the values for Warning and Fail.
b. For other fields, push the Edit softkey, use the arrow keys to edit the value, and
then push the Confirm softkey to save the value.
5. Once all threshold values are correct, push the Confirm softkey on the save the
threshold set.