8 Fluke Corporation Measuring uncommon RTDs with the Fluke 726
726 Serial Command List
Command Response/Actions Command Arguments Comment Ch 1 Ch 2
CUST1_ALIAS 1 argument, sets screen name for
CUST1 RTD.
CUST1_ALIAS? Returns screen name for CUST1 RTD Verify RTD 1 alias.
CUST2_ALIAS 1 argument, sets screen name for
CUST2 RTD.
CUST2_ALIAS? Returns screen name for CUST2 RTD. Verify RTD 2 alias.
CUST3_ALIAS1 1 argument, sets screen name for
CUST3 RTD.
CUST3_ALIAS? Returns screen name for CUST3 RTD. Verify RTD 3 alias
HART_ON Turns HART mode on. Switches in 250
ohm resistor.
HART_OFF Turns HART mode off. Switches out 250
ohm resistor.
HART? Returns state of hart mode, ON or OFF
•
*CLS Clear the error queue
•
FA U LT Returns error code FILO
•
ERROR CODES: NONNUMERIC_ENRTY (100)
EBUFFER_OVERFLOW (101)
INVALID_UNITS_CODE (102)
ENTRY_OVER_RNG (103)
ENTRY_UNDER_RNG (104)
MISSING_PARM (105)
INVALID_UNIT_PARM (106)
INVALID_SENSOR_TYPE (108)
UNKNOWN_COMMAND (110)
BAD_PARM_VALUE (111)
INPUT_BUFF_OVERFLOW (112)
MSG_BUFF_OVERFLOW (113)
OUTPUT_BUFF_OVERFLOW (114)
OUTPUT_OVERLOAD (115)
CAL_START Initiates a password protected See Cal. Manual
calibration (password = 627) for details
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2007 Fluke Corporation. All rights reserved.
Specifications subject to change without notice.
Printed in U.S.A. 6/2007 2546483 A-EN-N Rev B
Fluke. Keeping your world
up and running.
®