RPM4/HPMS A70M/A20M-AF Reference Pressure Monitor with High Pressure Mounting System Operation and Maintenance Manual

RPM4/HPMS A70M/A20M-AF OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company Page 6
2.1.2 INSPECTING CONTENTS
Check that all items are present and have no visible damage. If damage is noted, report it to
your Receiving Department for appropriate action.
A new RPM4/HPMS A70M/A20M-AF includes all items listed in Table 1.
Table 1. RPM4/HPMS A70M/A20M-AF Parts List
DESCRIPTION PART NO.
RPM4/HPMS A70M/A20M-AF Reference Pressure
Monitor/High Pressure Mounting System
402192
ACCESSORIES:
1 ea. Transport case (with inserts) (used for original
shipment)
124198
1 ea. Foot switch assembly 401613
1 ea. PGC-10000-AF Documentation Disk including:
RPM4/HPMS A70M/A20M-AF Operation and
Maintenance Manual, p/n 550136
GPC1-10000-AF Operation and Maintenance
Manual, p/n 550135
GB-152-AF Operation and Maintenance Manual,
p/n 550137
402189
1 ea. Calibration Report 550100
1 ea. Test Report (PGC-10000-AF) (if the RPM4/HPMS is
delivered as part of an PGC-10000-AF system)
550138
1 ea. Power cord (7.5 ft) 100770
1 ea. General Accessories Disk (white CD) (Important :
Includes system support software and
documentation)
102987
2.2 PGC-10000-AF SYSTEM
RPM4/HPMS A70M/A20M-AF is usually delivered as the reference pressure measurement component of
a PGC-10000-AF Pneumatic Gauge Calibrator. The PGC-10000-AF system includes:
RPM4/HPMS A70M/A20M-AF: Reference pressure monitor and high pressure mounting system used
as the pressure measuring reference of the calibration system.
GPC1-10000-AF: Gas pressure controller used to set and adjust high pressure gas in the calibration
system.
GB-152-AF: Gas booster package used to supply gas pressure up to 10 000 psi (70 MPa) to the
GPC1-10000-AF pressure controller.